Walker Filtration
Walker E30612MV Compressed Air Filter Element – Medical Vacuum (Grade MV)
- SKU:
- E30612MV
- MPN:
- E30612MV
Description
The Walker E30612MV is a genuine Walker Filtration medical vacuum replacement element for Alpha Medical Vacuum series filter housings. Deep-pleated media, tested in accordance with ISO 12500, delivers rated performance at a low differential pressure across the life of the element.
Medical Vacuum — Grade MV
Rated for >99.97% bacterial / viral removal. Protects medical vacuum plant and the people maintaining it, with greater than 99.97% retention of bacteria and viruses in line with NFPA 99, HTM 02-01 and ISO 7396-2.
Fits These Housings
Genuine Walker Filtration replacement element for the A30070MV, A30085MV and A30105MV housings. These housings are not stocked online — contact us for pricing and availability, or for help confirming which element your unit takes.
Replacement & Maintenance
Replace every 12 months, or sooner if the differential pressure indicator signals, to maintain rated performance. Genuine Walker Filtration element — part number E30612MV.
Additional Information
Grade: |
MV - Medical Vacuum |
Element Reference PN: |
E30612MV |
Particle / Aerosol Removal: |
>99.97% bacterial / viral |
Fits Housings: |
A30070MV, A30085MV, A30105MV |
Element Size: |
2.36" dia x 4.72" long |
Media: |
Deep-pleated borosilicate glass microfiber |
Tested To: |
ISO 12500 |